Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Assunto principal
Intervalo de ano de publicação
1.
Med Phys ; 23(1): 45-52, 1996 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-8700032

RESUMO

A device that generates low-energy x rays at the tip of a needle-like probe was developed for stereotactic interstitial radiosurgery. Electrons from a small thermionic gun are accelerated to a final energy of up to 40 keV and directed along a 3 mm outside diameter drift tube to a thin Au target, where the beam size is approximately 0.3 mm. All high-voltage electronics are in the probe housing, connected by low-voltage cable to a battery-operated control box. X-ray output, which is nearly isotropic, consists of a bremsstrahlung spectrum and several lines between 7 and 14 keV, with characteristic radiation contributing 15% of the total energy output. To date, 14 patients with metastatic brain tumors have been treated with this device.


Assuntos
Radiocirurgia/instrumentação , Fenômenos Biofísicos , Biofísica , Eletrônica Médica , Desenho de Equipamento , Humanos , Miniaturização , Tecnologia Radiológica
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...